Fig. 3From: Targeted thrombolysis by magnetoacoustic particles in photothrombotic stroke modelCharacterization of the acoustic system. (a) Schematic characterization of the full width at half maximum (FWHM). The acoustic pressure was driven from the linear amplifier connected with a function generator, and then the pressure was measured by hydrophone to convert the mechanical output into electrical signals. (b-c) FWHM of UT and FT, showing that a larger FWHM was detected in the UT than in the FT. (d) Schematic procedure of spatial characterization (Investigated size: 10 \(\times\) 10 mm\(^2\) for UT; 4 \(\times\) 4 mm\(^2\) for FT). (e) Evaluation of the focal diameter at -6 dB, indicating that 8.57 and 1.91 mm were measured in UT and FT, respectivelyBack to article page