Skip to main content

Table 5 ICP MS results of 4SEAL® Hemostatic powder

From: ISO 10993 biological evaluation of novel hemostatic powder – 4SEAL®

Analyzed element

Limit of detection (LOD) [μg/L]

Result [μg/L]

Total Element Exposure [μg]

Parenteral PDE [μg/day]

Cd

1

<  1

<  0.25

2

Pb

2.5

<  2.5

<  0.625

5

As

30

<  30

<  7.5

15

Hg

1

<  1

<  0.25

3

Co

2.5

<  2.5

<  0.625

5

V

24

<  24

<  6

10

Ni

10

<  10

<  2.5

20

Tl

5

<  5

<  1.25

8

Au

5

<  5

<  1.25

100

Pd

5

<  5

<  1.25

10

Ir

5

<  5

<  1.25

10

Os

5

<  5

<  1.25

10

Rh

5

<  5

<  1.25

10

Ru

5

<  5

<  1.25

10

Se

10

<  10

<  2.5

80

Ag

5

<  5

<  1.25

10

Pt

5

<  5

<  1.25

10

Li

10

<  10

<  2.5

250

Sb

10

<  10

<  2.5

90

Ba

10

<  10

<  2.5

700

Mo

10

<  10

<  2.5

1500

Cu

130

<  130

<  32.5

300

Sn

10

<  10

<  2.5

600

Cr

60

<  60

<  15

1100