Skip to main content

Table 8 AMES overall results

From: ISO 10993 biological evaluation of novel hemostatic powder – 4SEAL®

Strain

Mutagenic data points

Overall Result for sample

Solvent control

Positive control

 

w/o S9

with S9

 

w/o S9

with S9

w/o S9

with S9

TA98

No

No

Probably not mutagenic

PASS

PASS

PASS

PASS

TA100

No

No

Probably not mutagenic

PASS

PASS

PASS

PASS

TA1535

No

Yes

Unclear, needs further evaluation

PASS

PASS

PASS

PASS

TA1537

No

No

Probably not mutagenic

PASS

PASS

PASS

PASS

E. coli uvrA [pKM101]

No

No

Probably not mutagenic

PASS

PASS

PASS

PASS