From: Effect of sterilization on 3-point dynamic response to in vitro bending of an Mg implant
Test feature or detail. | Technical data. | Observations. |
---|---|---|
Distance between centers L. | 90 mm. | - |
Test speed (velocity). | 5 mm/min. | F1264-03. |
Used machine. | Instron 1011, capacity 5 kN. | January 25, 2019 calibration (Mess. Serv. Metrological). |
Load amplitude applied. | 555 N. | 25% Máximum permissible load. |
Frequency. | 5 Hrz. | Load ratio ≥ 0.1 |
Cross section of the specimen. | 71.48 mm 2. | - |
Specimen length. | 114 mm. | - |
Specimen material. | WE43 Casting bars. | Y 3.7-4.3, Zr 0.4-1.0, Gd 0-1.9, Nd 2.0-2.5, Cu ≤ 0.02, Ni ≤ 0.005, Fe ≤0.01, Mg balance |
Average weight specimen. | 0.01511 kg. | - |
Test temperature. | 27 °C. | - |